Echo 555 LIQUID HANDLER

Echo 555 LIQUID HANDLER

for high throughput screening

  • Fastest transfer speeds of the Echo Liquid Handler family
  • Broad application range
  • Fast, accurate, and precise liquid transfers in nanoliter volumes
REQUEST INFO

Addressing the Demands of High-Throughput Screening Labs

The Echo 555 Liquid Handler has the speed and flexibility required for a screening lab. With rapid, highly precise nL volume transfers, Echo 555 Liquid Handler can process over 650* compound and over 500* reagent plates per day. Tasks like plate reformatting, hit picking, daughter plate creation, and adding samples and reagents for screening assays are made easy with the Echo 555 Liquid Handler. *Based on 100 nL volume transfers

Key Benefits

  • Tipless transfer, resulting in dramatic cost savings
  • Improved information about sample integrity and improved transfer quality through measurement of DMSO hydration level
  • Elimination of compound loss, cross-contamination, carryover or leachates through non-contact transfer
  • Extend compound library life through miniaturization of assays
  • Greatly simplified assay development and dilution processes through the flexibility to transfer any volume of liquid from any well to any other well

A Broad Range of Applications

Going beyond DMSO used in high throughput screening, the Echo 555 system supports the transfer of a broad range of aqueous solutions to support high throughput screening laboratories.

icon-drug-discovery-app

for DRUG DISCOVERY

  • HIGH-THROUGHPUT SCREENING
  • HIGH CONTENT SCREENING
  • CONCENTRATION CURVE GENERATION

Echo Acoustic Liquid Handling|by Labcyte Inc.

Watch the Echo acoustic droplet ejection (ADE) technology in action. Sound waves eject precisely-sized droplets from the source plate onto the destination plate suspended above.

Impacting Therapeutic Discovery

Tipless, non-contact, acoustic fluid transfer of the Echo 555 Liquid Handler eliminates the potential for samples or compounds to adhere to tips. This helps to improve data quality, increasing the potential to find new therapeutics.

In a screen of 975 compounds, Bristol-Myers Squibb found 108 more hits when using the Echo Liquid Handler1.

1 Spicer T., Fitzgerald Y., Burford N., Matson S., Chatterjee M., Gilchrist M.,Myslik J., O'Connell J.
Pharmacological evaluation of different compound dilution and transfer paradigms on an enzyme assay in low volume 384-well format. Poster presented at Drug Discovery Technology, August 2005, Boston, MA.

32% More Active Compounds Found

Potent by

Echo transfer

108 | 32%

Potent by

both methods

230 | 68%

icon-maximize-budget

Save Compounds and Money by Miniaturizing Assays

Utilize the Echo 555 Liquid Handler to set up 384-well cell-based and biochemical assays or transition into 1536-well formats with no loss of performance. The nL volume transfers of the Echo 555 systems enable direct dilution of concentrated compounds – eliminating workflow steps and keeping the final concentration of DMSO low in the final assay. Data accuracy is further improved through the elimination of compound loss on tips.

FIGURE 1: Stimulation curves and EC50 values for isoproterenol, salbutamol, and salmeterol in 384-well (left) and 1536-well (right), LV white microplate. Incubation time of 60 minutes. BMG Labtech PHERAstar FS.1

  • Isoprotenol
  • Salbutamol
  • Salmeterol

FIGURE 1: Stimulation curves and EC50 values for isoproterenol, salbutamol, and salmeterol in 384-well, LV white microplate. Incubation time of 60 minutes. BMG Labtech PHERAstar FS.1

Description EC50
Isoproterenol 0.8 nM
Salbutamol 11 nM
Salmeterol 3.2 nM
Z’ Factor 0.72
  • Isoprotenol
  • Salbutamol
  • Salmeterol

FIGURE 2: Stimulation curves and EC50 values for isoproterenol, salbutamol, and salmeterol in 1536-well, LV white microplate. Incubation time of 60 minutes. BMG Labtech PHERAstar FS.1

Description EC50
Isoproterenol 0.8 nM
Salbutamol 27 nM
Salmeterol 0.2 nM
Z’ Factor 0.89

1Edwards B., Tang N., Nakamura M. Miniaturized HTRF Assays Using the Echo Liquid Handler. Labcyte Application Note.

Icon Echo CS

Making High-throughput Combination Screening a Reality

Drug combination screening using the Echo system has revolutionized the approach to the assessment of synergistic effects of drug combinations on disease. The capability to transfer from any-well-to-any-well and the Echo Combination Screening Software make high throughput, 2-way and 3-way dose response matrix studies a reality.

Echo 555 Specifications

Description Specifications
Drop volume 2.5 nL
Volume transfer range 2.5 nL – 10 µL (dependent on fluid type, assay, and source plate used)
Reagents supported DMSO, PCR and qPCR reagents, synthetic biology reagents, NGS reagents, primers and probes, cell culture media, proteins, nucleic acids, up to 50% glycerol, serum, plasma, antibodies, enzymes, and protein crystallography reagents
Transfer accuracy <10% deviation from target volume
Transfer precision <8% CV (except protein crystallography fluid class)
Source labware compatibility All Echo Qualified 384- and 1536-well Microplates
Destination labware compatibility All Echo Qualified Microplates and most ANSI-compliant/SBS-standard microplates in 96-, 384-, 1536-, and 3456-well formats 8 to 16 mm well height
Barcode reader options Short-side and/or long-side
Dimensions 53.9 cm width x 68.3 cm depth x 92.5 cm height (21.2 x 26.9 x 36.4 inches)
Weight 128 kg (283 lbs)
Working envelope Additional 25.4 cm (10 inches) on top, 2.5 cm (1 inch) on sides, 7.6 cm (3 inches) at front, 30.5 cm (12 inches) at back
Electrical AC 120V, 50/60 Hz, 10 A or AC 230V, 50/60Hz, 5 A
Operating temperature 21°C ± 5°C (70°F ± 9°F)
Vacuum House supply minimum 200 Torr (266 mbar, 22 inches Hg Vac)
OR
Standalone vacuum with surge tank, minimum pump speed 2.0 m3/hr, minimum pressure 100 Torr (133 mbar, 26 inches Hg Vac)
Fluid class calibration No calibration needed